14.4V Smart Battery Pack Solutions for Medical Ventilators
Uninterrupted power for life-support equipment — without a custom battery development cycle.
A ventilator cannot tolerate a power gap. When AC mains fails or a patient moves between wards, the battery pack is the only power source standing between the device and a stopped turbine. TEFOO’s standard 14.4V SMBus smart battery packs give ventilator manufacturers a pre-certified, in-stock power platform: four models covering 47.52Wh to 95.04Wh, continuous discharge up to 15.0A, hot-swap-ready fuel-gauge data, and IEC 62133-2 / UN 38.3 documentation already in place.
Contact our engineering team →
Ventilator Power Requirements the Battery Pack Must Cover
Continuous and Peak Current
A ventilator’s battery load is set by three subsystems: the turbine/blower (dominant load, with a startup surge as the peak current event), control electronics and display (constant baseline), and an optional heated humidifier (which can double the total power budget). Pack selection must be based on worst-case continuous draw plus margin — not the nominal operating point.
|
Load Type |
Typical Range |
Sizing Requirement |
|---|---|---|
|
Continuous draw |
10–40W (0.7–2.8A @ 14.4V) |
Pack max discharge ≥ 1.3× device max continuous draw |
|
Peak draw (turbine spin-up, valve actuation) |
1.5–2× continuous, duration <1s |
Protection circuit must not trip on inrush |
|
Worst case with humidifier |
Up to ~50–60W |
Use 8.0A (GS2054HH) or 15.0A (GS2024HH) models |
The four 14.4V models span 4.0A to 15.0A max discharge current, so the pack matches the device class without over-specifying weight.
Hot-Swap and Dual-Battery Operation
For patients who cannot tolerate ventilation interruption, dual-battery hot-swap architecture is the standard design answer: the device draws from one pack while the depleted pack is replaced, with no power gap. This architecture depends on accurate, independent data from every pack — which is what the SMBus v1.1 interface on all TEFOO standard packs provides:
- Per-pack fuel gauge data: State of Charge, remaining capacity, cycle count, and temperature reported over SMBus v1.1, with fuel gauge accuracy up to 1%
- LED RSOC indicator: remaining charge visible at a glance, without powering on the device
- Power path management: the GSPMB20 battery power management module handles boost/buck charging and seamless switching between mains and battery power over SMBus; multiple GSPMB20 units support parallel powering of battery packs
- Mating hardware: GSC550M18012D / GSC550M09010D connectors and the GSCBB connector board are available for host-side battery bay design
Standard 14.4V Models for Ventilator Applications
Four standard models form two energy tiers — 47.52Wh (GS2034DH, GS2054DH) and 95.04Wh (GS2054HH, GS2024HH) — all sharing the same SMBus v1.1 interface and +/C/D/T/- pinout, so one battery bay design supports multiple energy options.
Technical Specifications
|
Parameter |
||||
|---|---|---|---|---|
|
Configuration |
4S1P |
4S1P |
4S2P |
4S2P |
|
Rated voltage |
14.4V |
14.4V |
14.4V |
14.4V |
|
Rated capacity |
3.3Ah (nominal 3.45Ah) |
3.3Ah (nominal 3.45Ah) |
6.6Ah (nominal 6.9Ah) |
6.6Ah (nominal 6.9Ah) |
|
Energy |
47.52Wh |
47.52Wh |
95.04Wh |
95.04Wh |
|
Max discharge power |
60W |
60W |
100W |
200W |
|
Max discharge current |
4.0A |
4.0A |
8.0A |
15.0A |
|
Max charging current |
1.65A |
1.65A |
4.83A |
4.83A |
|
Max charging voltage |
16.8V |
16.8V |
16.8V |
16.8V |
|
Internal resistance |
<200mΩ |
<200mΩ |
<150mΩ |
<100mΩ |
|
Dimensions |
150.5×42.0×22.4mm (5.93×1.65×0.88in) |
85.1×77.4×22.7mm (3.35×3.05×0.89in) |
150.5×77.4×22.7mm (5.93×3.05×0.89in) |
167.3×106.8×21.4mm (6.59×4.20×0.84in) |
|
Weight |
240g ±10g (8.47oz) |
235g ±10g (8.29oz) |
445g ±10g (15.70oz) |
500g ±10g (17.64oz) |
|
Charging temperature |
0°C~50°C (32°F~122°F) |
0°C~50°C (32°F~122°F) |
0°C~50°C (32°F~122°F) |
0°C~50°C (32°F~122°F) |
|
Discharge temperature |
-20°C~60°C (-4°F~140°F) |
-20°C~60°C (-4°F~140°F) |
-20°C~60°C (-4°F~140°F) |
-20°C~60°C (-4°F~140°F) |
|
Cycle life |
500 cycles, ≥80% |
500 cycles, ≥80% |
500 cycles, ≥80% |
500 cycles, ≥80% |
|
Interface |
+/C/D/T/-, SMBus v1.1 |
+/C/D/T/-, SMBus v1.1 |
+/C/D/T/-, SMBus v1.1 |
+/C/D/T/-, SMBus v1.1 |
All four models: SMBus v1.1 Smart Battery specification · integrated LED RSOC indicator · comprehensive protection functions · integrated cell balancing · fuel gauge accuracy up to 1%.
Model Selection Guide
|
Selection Dimension |
GS2034DH |
GS2054DH |
GS2054HH |
GS2024HH |
|---|---|---|---|---|
|
Best-fit device class |
Slim portable ventilators |
Compact portable ventilators |
Transport ventilators |
ICU / high-flow platforms |
|
Single-pack runtime @ 25W |
~1.9h |
~1.9h |
~3.8h |
~3.8h |
|
Dual-pack hot-swap runtime @ 25W |
~3.8h |
~3.8h |
~7.6h |
~7.6h |
|
Peak load headroom |
Moderate (4.0A) |
Moderate (4.0A) |
High (8.0A) |
Very high (15.0A) |
|
Weight impact on device |
Lowest in class |
Lowest in class |
+205g vs DH models |
+265g vs DH models |
|
MOQ |
None |
None |
None |
None |
|
Standard lead time |
7–30 days |
7–30 days |
7–30 days |
7–30 days |
Selection rule: choose the smallest pack that meets your runtime and peak-current requirements. If none of the four standard envelopes fits your mechanical or electrical constraints, TEFOO provides customized battery pack solutions with tailored mechanical, electrical, and firmware parameters.
Runtime Calculation
Runtime (hours) = usable battery energy (Wh) ÷ average device power consumption (W)
|
Average Device Power |
GS2034DH / GS2054DH (47.52Wh) |
GS2054HH / GS2024HH (95.04Wh) |
Dual 95.04Wh (hot-swap) |
|---|---|---|---|
|
15W |
3.17h |
6.34h |
12.67h |
|
25W |
1.90h |
3.80h |
7.60h |
|
35W |
1.36h |
2.72h |
5.43h |
|
50W |
0.95h |
1.90h |
3.80h |
The power levels above are illustrative calculation rows — substitute your measured average power into the formula for an exact figure. Runtimes use nominal energy with no derating applied.
For field planning, apply three corrections to the nominal figure:
- Measure real average power across the full breathing cycle at the clinical settings in use — nameplate power overestimates, idle-state power underestimates
- Derate for aging: cycle life is specified at ≥80% capacity retention after 500 cycles; plan fleet replacement against end-of-life capacity
- Derate for temperature: discharge capacity at -20°C is lower than at 25°C; size transport and emergency-use devices against the coldest expected operating point within the -20°C~60°C discharge range
A conservative engineering practice is to plan against 80–90% of nominal energy.
Compliance and Certifications
Three standards operate at different levels of the ventilator supply chain, and the documentation for the battery level is already in place:
- IEC 60601-1 (medical electrical equipment): certifies the ventilator as a complete device. A battery pack with certified pack-level safety (IEC 62133-2 / UL2054), accurate SMBus fuel-gauge data, and predictable protection behavior simplifies the host device’s 60601-1 evaluation — particularly the clauses covering power source interruption, battery monitoring, and single-fault conditions.
- IEC 62133-2 / UL2054 (battery safety): the standard packs are certified to IEC 62133-2 and UL2054, covering safety requirements for portable sealed lithium-ion cells and battery packs. These reports transfer directly into your device technical file.
- UN 38.3 (transport): all packs have passed UN 38.3 testing with MSDS documentation — the prerequisite for legal air and sea freight of lithium batteries. No additional transport testing is needed before shipping.
- Regional marks: CE, FCC, RoHS, UKCA, PSE, and KC are in place; additional market-specific certificates can be arranged per customer requirements.
- Manufacturer quality systems: TEFOO (Shenzhen GRACE Technology Development Co., Ltd.) operates under ISO 13485 (medical device quality management), ISO 9001, ISO 14001, and ISO 45001 certified systems.
For supplier qualification, request the IEC 62133-2 / UL2054 test reports and the UN 38.3 test summary — together with the manufacturer’s ISO 13485 system, they form the battery-related core of a ventilator’s technical file.
Lead Time, MOQ, and Supply Continuity
Medical device development cycles are long and complex. Standardized smart battery packs based on SMBus communication remove the battery system from that critical path:
- No battery system development — electrical interface, fuel gauge, and protection behavior are fixed and documented
- Faster product validation and certification — existing IEC 62133-2 / UL2054 and UN 38.3 reports enter your submission directly
- Reduced engineering workload — one SMBus v1.1 interface covers all four models
- Accelerated time-to-market — standard products are held in stock, with a 7–30 day lead time
Medical devices remain in service for many years, so supply continuity matters as much as first-build cost. TEFOO supports long-term availability with consistent battery specifications, long-term supply capability, in-stock standard products, and no minimum order quantity (MOQ) — covering everything from prototype builds to spare and replacement demand across the full service life of your device. Custom projects follow a project-specific timeline.
Power Your Ventilator with a Proven Battery Platform
Standard models ship in 7–30 days with no MOQ. For custom mechanical, electrical, or firmware requirements — or additional market-specific certifications — our engineering team responds with a solution proposal.
Request a quote or technical consultation →
Manufacturer: Shenzhen GRACE Technology Development Co., Ltd. | www.tefoobattery.com